Volume 23, Issue 9 (Nov-Dec 2015)                   JSSU 2015, 23(9): 806-818 | Back to browse issues page

XML Persian Abstract Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Ahmadi O, Tavakoli-Anbaran H. Calculating Error Percentage in Using Water Phantom Instead of Soft Tissue Concerning 103Pd Brachytherapy Source Distribution via Monte Carlo Method. JSSU 2015; 23 (9) :806-818
URL: http://jssu.ssu.ac.ir/article-1-2831-en.html
Abstract:   (6061 Views)

Introduction: 103Pd is a low energy source, which is used in brachytherapy. According to the standards of American Association of Physicists in Medicine, dosimetric parameters determination of brachytherapy sources before the clinical application was considered significantly important. Therfore, the present study aimed to compare the dosimetric parameters of the target source using the water phantom and soft tissue.

Methods: According to the TG-43U1 protocol, the dosimetric parameters were compared around the 103Pd source in regard with water phantom with the density of 0.998 gr/cm3 and the soft tissue with the density of 1.04 gr/cm3 on the longitudinal and transverse axes using the MCNP4C code and the relative differences were compared between the both conditions.

Results: The simulation results indicated that the dosimetric parameters depended on the radial dose function and the anisotropy function in the application of the water phantom instead of soft tissue up to a distance of 1.5 cm,  between which a good consistency was observed. With increasing the distance, the difference increased, so as within 6 cm from the source, this difference increased to 4%.

Conclusions: The results of  the soft tissue phantom compared with those of the water phantom indicated 4% relative difference at a distance of 6 cm from the source. Therefore, the results of the water phantom with a maximum error of 4% can be used in practical applications instead of soft tissue. Moreover, the amount of differences obtained in each distance regarding using the soft tissue phantom could be corrected.

Full-Text [PDF 625 kb]   (1831 Downloads)    
Type of Study: Original article | Subject: Medical Physics
Received: 2014/06/8 | Accepted: 2015/11/17 | Published: 2015/12/28

Add your comments about this article : Your username or Email:
CAPTCHA

Send email to the article author


Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.

© 2024 CC BY-NC 4.0 | SSU_Journals

Designed & Developed by : Yektaweb